• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COLOPLAST A/S PERISTEEN TRANSANAL IRRIGATION SYSTEM WITH CATHETER; PERISTEEN PLUS SYSTEM R. - T.B

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S PERISTEEN TRANSANAL IRRIGATION SYSTEM WITH CATHETER; PERISTEEN PLUS SYSTEM R. - T.B Back to Search Results
Model Number 2914001005
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abdominal Pain (1685); Diarrhea (1811); Ischemia (1942); Vomiting (2144)
Event Date 09/15/2022
Event Type  Injury  
Event Description
According to the available information, coloplast was notified of the incident that occurred after use of the peristeen plus transanal irrigation system.It was reported that on (b)(6) 2022 the user performed regular procedure with use of the device without any issues.Evacuation started and then suddenly stopped.The user felt a major (¿huge¿) pain.The user went to bed without eating, the pain continued, she vomited and in the evening, she went to the hospital ((b)(6)).The morning after she was hospitalized and operated on.According to the available information, the patient did not sustain perforation.No further information is available at this time.
 
Manufacturer Narrative
Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Manufacturer Narrative
Medical assessment has been added to the "report incident description" field."report incident description field" has been updated.
 
Event Description
According to the information available, this case concerns a 45-year-old woman suffering from cauda equina and multiple sclerosis, which over time prompted a neurogenic bladder and bowel.In 1996 she had a right sided adnexectomy.The patient irrigated since 2005 (enemas) every 2nd day, using peristeen since 2011 without essential problems, apart from ¿incomplete evacuation¿.On the day of the incident the patient irrigated as normal when the evacuation procedure suddenly stopped and strong abdominal pain developed.At the emergency department a mechanical adherence ileus (string ileus) at the site of the previous surgery was diagnosed and confirmed at the subsequent laparoscopy.The affected 30 cm long lieal segment, located 40 cm from the ileocecal valve, was resected, and a vital anastomosis established.Eventless recovery, the patient was discharged from hospital day 7.No further information was provided.
 
Manufacturer Narrative
After further evaluation, it has been justified that based on the medical assessment, this incident was not device related.A causal relationship between peristeen transanal irrigation and a mechanical ileus of an ileal segment in the right iliac fossa is unlikely and the incident is considered as coincidental.Therefore, this incident is no longer considered reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERISTEEN TRANSANAL IRRIGATION SYSTEM WITH CATHETER
Type of Device
PERISTEEN PLUS SYSTEM R. - T.B
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebæk 3050
DA  3050
Manufacturer (Section G)
COLOPLAST HUNGARY - NYI
coloplast utca 2
nyírbátor 4300
HU   4300
Manufacturer Contact
usmaco maria bosse
1601 west river road north
minneapolis, MN 55411
8007880293
MDR Report Key16077756
MDR Text Key306424793
Report Number3006606901-2022-00024
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
PMA/PMN Number
K140310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number2914001005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/05/2022
Initial Date FDA Received12/30/2022
Supplement Dates Manufacturer Received12/05/2022
12/05/2022
Supplement Dates FDA Received01/13/2023
02/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention; Life Threatening;
Patient Age45 YR
Patient SexFemale
Patient Weight60 KG
-
-