• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY Back to Search Results
Catalog Number PXSLIMLAN135STR
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2022
Event Type  malfunction  
Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event Description
The patient was undergoing a coil embolization procedure in the superficial femoral artery (sfa) using a lantern delivery microcatheter (lantern) and a non-penumbra catheter.After completing the procedure, the physician fractured the lantern mid-shaft while removing it from the patient.Therefore, the based catheter containing the fractured lantern were removed together as unit.The procedure was considered completed.There was no report of an adverse effect to the patient.
 
Manufacturer Narrative
Evaluation of the returned lantern confirmed a fractured and revealed kinks near the fracture.The kinks near the fracture indicate that the device likely kinked prior to fracturing.If the device is manipulated against resistance or is otherwise mishandled during retraction, damage such as a kink and subsequent fracture may occur.Further evaluation revealed an ovalization on the distal end of the microcatheter.This damage was likely incidental to the reported complaint and the root cause could not be determined.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN DELIVERY MICROCATHETER
Type of Device
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key16080895
MDR Text Key307039022
Report Number3005168196-2022-00611
Device Sequence Number1
Product Code DQY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152840
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/01/2005,02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPXSLIMLAN135STR
Device Lot NumberF00002375
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 12/07/2022
Initial Date FDA Received12/30/2022
Supplement Dates Manufacturer Received02/09/2023
Supplement Dates FDA Received02/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-