• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616
Device Problems Failure to Shut Off (2939); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2022
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that the motor drive unit was stuck in forward and running without activation.It is unknown whether the event happened during surgery and if there was patient involvement and if there was a back-up device available or if there was a delay.No further complications were reported.
 
Manufacturer Narrative
H10: internal complaint reference (b)(4).
 
Event Description
It was reported that during set up, the motor drive unit was stuck in forward and running without activation.A back-up device was available and no surgical delay was reported.There was no patient involvement.
 
Manufacturer Narrative
H10: internal complaint reference case-(b)(4).H3, h6: the reported device was received for evaluation.There was no relationship found between the device and the reported event.A visual inspection found a discolored cable.A functional evaluation was performed on the returned device and found failure to shut off and physical resistance were not noted during testing a review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Factors that could have contributed to the reported event include a failure of a concomitant device.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR DRIVE UNIT, HAND CNTRL, PWRMX EL
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16082667
MDR Text Key306521309
Report Number1643264-2023-00001
Device Sequence Number1
Product Code HAB
UDI-Device Identifier03596010555472
UDI-Public03596010555472
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200616
Device Catalogue Number72200616
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/14/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-