• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 DOT 3FR SL PROV ML; MIDLINE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 DOT 3FR SL PROV ML; MIDLINE CATHETER Back to Search Results
Model Number N/A
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/05/2022
Event Type  malfunction  
Event Description
T was reported by the customer "the micro introducer sheath accordioned or folded back upon insertion through the skin.Nurse attempted to push pass the skin with minimal success and then tried to pull it out after slightly penetrating the skin and then felt a snug removal.Device was removed successfully with no issue.Additional information received 12/16/2022: it was reported by the healthcare professional ¿i was set up to place the midline.Due to issues, we had been having i thoroughly inspected the introducer before insertion.Everything looked normal.I do not nick before inserting as most insert without issue.I was meeting resistance, so i pulled the introducer back, made a slight nick and reinserted.I did not inspect the introducer before trying the second time, but as it entered the epidermis and then dermis, i was feeling a lot of resistance, like i was going through the back of carpet.I pulled the catheter out and that is when i noticed that the grey part of the introducer had two frays.Another kit (picc kit) was opened, and the introducer used out of that kit with no issues.Device was secured with a statlock.Catheter is still intact due to documentation.No harm but additional pressure/pain during insertion and reinsertion of second introducer.
 
Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of regp1017 showed four other similar product complaint(s) from this lot number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOT 3FR SL PROV ML
Type of Device
MIDLINE CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key16088217
MDR Text Key308493538
Report Number3006260740-2022-05961
Device Sequence Number1
Product Code PND
UDI-Device Identifier00801741185076
UDI-Public(01)00801741185076
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Nurse
Type of Report Initial
Report Date 12/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberSP4153108DB
Device Lot NumberREGP1017
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2022
Initial Date FDA Received01/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/01/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age50 YR
Patient SexFemale
Patient Weight73 KG
-
-