• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL G4 STEERABLE GUIDING CATHETER; CATHETER, STEERABLE Back to Search Results
Model Number SGC0701
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Low Blood Pressure/ Hypotension (1914); Pericardial Effusion (3271); Unspecified Tissue Injury (4559)
Event Date 12/14/2022
Event Type  Injury  
Manufacturer Narrative
The device will not be returned for evaluation as the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report tissue damage and pericardial effusion.It was reported this was a mitraclip procedure to treat functional mitral regurgitation (mr) with a grade of 4 and a pericardial effusion.The steerable guide catheter (sgc) was inserted, but the effusion had slightly changed.The physician was unable to determine if the sgc worsened the effusion.It was then observed the patients blood pressure the decreases and a tissue flap was observed on the septum.Therefore, the sgc was removed, and the procedure was discontinued.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.The investigation was unable to determine a cause for the reported tissue injury and increased of pericardial effusion cannot be determined.The hypotension appears to be a cascading effect of the pericardial effusion.Tissue injury, pericardial effusion, and hypotension are listed in the instructions for use as known possible complications associated with mitraclip procedures.The reported serious injury/illness/impairment was a result of case specific circumstance as no treatment was provided for the patient effects.There is no indication of a product issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G4 STEERABLE GUIDING CATHETER
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16091745
MDR Text Key306591817
Report Number2135147-2023-00016
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648231025
UDI-Public08717648231025
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/11/2023
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number20412R336
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/14/2022
Initial Date FDA Received01/03/2023
Supplement Dates Manufacturer Received01/27/2023
Supplement Dates FDA Received02/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/12/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age84 YR
Patient SexFemale
-
-