Catalog Number 5F060801C |
Device Problems
Break (1069); Positioning Failure (1158); Fracture (1260); Misfire (2532); Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a stent placement procedure, stent allegedly partially deployed.It was further reported that stent allegedly broke out of the holder before it could be released.Reportedly the stent was removed.There was no reported patient injury.
|
|
Manufacturer Narrative
|
The catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent are identified in d2 and g4.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Device pending return.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the lifestent vascular stent that are cleared in the us.The pro code and 510 k number for the lifestent vascular stent are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the sample was returned for evaluation.The returned delivery system was found in used condition with the loaded stent; a force transmitting joint was found broken in the catheter section which made a deployment impossible.The inner catheter cardan tube was found broken and separated in the mid-section which was considered a cascading event.Based on the information available the investigation is closed with confirmed result for break and separation of a force transmitting joint and subsequent break of the inner catheter cardan tube.A definite root cause for the reported event could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use states: 'do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit.' regarding pre dilation the instructions for use states: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' under materials required the instructions for use states: '5f (1.67 mm) or larger introducer sheath (¿); 0.014-inch (0.36 mm) - 0.035-inch (0.89 mm) diameter guidewire'.Holding and handling of the system throughout deployment was found sufficiently described.H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a stent placement procedure, stent allegedly partially deployed.It was further reported that stent allegedly broke out of the holder before it could be released.Reportedly the stent was removed.There was no reported patient injury.
|
|