• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT QUADRA ASSURA MP ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3371-40QC
Device Problems Intermittent Capture (1080); Over-Sensing (1438)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2022
Event Type  malfunction  
Manufacturer Narrative
Session records were provided for review by technical services.Analysis of the session records indicated that the pptwos event was sensed by the device, due to intermittent loss of capture on the left ventricular channel.
 
Event Description
It was reported that during a follow up in clinic, post-paced t-wave oversensing (pptwos) was noted on the device.Abbott technical support was contacted and the pptwos was suspected to be due to intermittent loss of capture on the left ventricular channel.Reprogramming of the device was recommended, however, no intervention has been performed at this time.The patient was in stable condition and will continue to be monitored.
 
Event Description
Additional information was received indicating the device was reprogrammed to resolve the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16098577
MDR Text Key307290224
Report Number2017865-2023-00330
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Model NumberCD3371-40QC
Device Lot NumberP000085459
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/15/2022
Initial Date FDA Received01/04/2023
Supplement Dates Manufacturer Received04/27/2023
Supplement Dates FDA Received04/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; QUARTET LEAD SMALL-S, 86 CM; TENDRIL STS
-
-