ZIMMER BIOMET, INC. DISTAL MEDIAL HUMERAL PLATE SHORT LEFT; PROSTHESIS, SHOULDER Back to Search Results
Model Number 47-2358-008-05
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Patient Problems Failure of Implant (1924); Pain (1994); Non-union Bone Fracture (2369)
Event Date 10/21/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: unk zplp 3.5mm locking screws.Unk zplp 3.5mm locking screws.Unk zplp 3.5mm locking screws.Unk zplp 3.5mm locking screws.Report source: foreign- united kingdom.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device has been discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2023 - 00029, 0001822565 - 2023 - 00030, 0001822565 - 2023 - 00031, 0001822565 - 2023 - 00032.
 
Event Description
It was reported during a retrospective clinical study that the patient was diagnosed radiographically with a non-union.Subsequently, the patient was revised five years post initial implantation for removal of implants and a bone graft was done.No additional patient consequences were reported.
 
Manufacturer Narrative
(b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was further reported that the patient had a orif of the left distal humerus.Subsequently, approximately 7 years ago was revised due to increasing pain, struggle with end range extension and evidence of loosening of the proximal screws and plate, and malunion at the fracture site found on radiographic imaging.During the revision, a bone graft was placed at the malunion site and all components were removed and a new distal-medial humeral plate with k-wires was implanted.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).Reported event was confirmed as review of the available records identified a orif distal humeral fracture complicated by proximal screw loosening and fracture non-union.Subsequent orif revision with subsequent plate retraction and extensive screw loosening with persistent non-union.Extensive screw loosening of both the original humeral orif and subsequent revision with fracture non-union and malalignment as described.Initial bone quality was mildly osteopenic with marked progression over time to severe osteopenia.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).D10 - medical product: catalog #: unknown, zplp 3.5mm locking screws, lot # unknown qty 4.Catalog #: 47235800609, distal posterior/lateral humeral plate, lot # 61208368.Catalog #: 47235901238, 3.5 mm locking screw with 2.7 mm head 12 mm length, lot # 61117349.Catalog #: 47482801202, 2.7 mm locking screw 12 mm length, lot # 60625467.Catalog #: 47482802202, 2.7 mm locking screw 22 mm length, lot # 60646645.Catalog #: 47482803802, 2.7 mm locking screw 38 mm length, lot # 61126468.Catalog #: unknown, cortical bone screw self-tapping hex head, lot # 60837245.Catalog #: 47235902438, 3.5 mm locking screw with 2.7 mm head 24 mm length, lot # unknown catalog #: unknown, unk screw, lot # unknown.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-01794 , 0001822565-2023-01795, 0001822565-2023-01796, 0001822565-2023-01797, 0001822565-2023-01798, 0001822565-2023-01799, 0001822565-2023-01800, 0001822565-2023-01894.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause cannot be determined based on additional information received.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISTAL MEDIAL HUMERAL PLATE SHORT LEFT
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16102180
MDR Text Key306697733
Report Number0001822565-2023-00025
Device Sequence Number1
Product Code HRS
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K082078
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Study,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Model Number47-2358-008-05
Device Catalogue Number47235810807
Device Lot Number61153872
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/22/2022
Initial Date FDA Received01/05/2023
Supplement Dates Manufacturer Received01/30/2023
05/02/2023
06/15/2023
01/19/2024
Supplement Dates FDA Received02/16/2023
05/04/2023
07/12/2023
01/26/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/12/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age87 YR
Patient SexFemale
Patient Weight86 KG
-
-