ZIMMER BIOMET SPINE INC. PEDICLE PROBE - CURVED LUMBAR LENKE Back to Search Results
Catalog Number 07.02072.001
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2022
Event Type  malfunction  
Event Description
It was reported that a vitality probe broke intra-op; the tip was recovered using pliers.There was a delay of 5 minutes without patient impacts.
 
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Manufacturer Narrative
Device evaluation: visual inspection revealed that the tip of lot 112489 was fractured off.Potential cause: root cause was unable to be determined.This event could possibly be attributed to wear through use over time or multiple sterilization cycles.It could also be attributed to off-axis forces applied during use.Dhr review and related actions: per dhr review, the parts were likely conforming when they left zimvie control.Device usage: this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
 
Event Description
It was reported that a vitality probe broke intra-op; the tip was recovered using pliers.There was a delay of 5 minutes without patient impacts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEDICLE PROBE - CURVED LUMBAR LENKE
Type of Device
PROBE
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
kim martinez
10225 westmoor dr.
na
westminster, CO 80021
3035144809
MDR Report Key16102563
MDR Text Key306794850
Report Number3012447612-2023-00011
Device Sequence Number1
Product Code HXB
UDI-Device Identifier00889024009943
UDI-Public(01)00889024009943(10)112489
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07.02072.001
Device Lot Number112489
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/29/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-