• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Model Number ER420
Device Problem Firing Problem (4011)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # x95x0w.Investigation summary : the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the er420 device was returned with a clip in the jaws and with no apparent damage.In addition, the packaging opened was returned along with the instrument.The clip was removed in order to inspect the jaws and they were found with no damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the trigger could not be activated due to be jammed.The device was disassembled to evaluate the condition of the internal components and the trigger was found bent, not allowing the clips to fed into the jaws.In addition, 11 clips were found remaining inside the clip track.The damaged on the device could be that the internal ribs were being gauged by the feeder link, causing the trigger to experienced additional force that could interfere with the firing of the device.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot 881a33 number, and no non-conformances were identified.
 
Event Description
It was reported that during a laparoscopic gastric bypass, the firing trigger on the er420 clip applier was sticking.Case completed with a like device.There was no patient harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key16106223
MDR Text Key308542284
Report Number3005075853-2023-00097
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012597
UDI-Public10705036012597
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberER420
Device Catalogue NumberER420
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2022
Initial Date Manufacturer Received 12/15/2022
Initial Date FDA Received01/05/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-