(b)(4).Batch # x95x0w.Investigation summary : the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the er420 device was returned with a clip in the jaws and with no apparent damage.In addition, the packaging opened was returned along with the instrument.The clip was removed in order to inspect the jaws and they were found with no damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the trigger could not be activated due to be jammed.The device was disassembled to evaluate the condition of the internal components and the trigger was found bent, not allowing the clips to fed into the jaws.In addition, 11 clips were found remaining inside the clip track.The damaged on the device could be that the internal ribs were being gauged by the feeder link, causing the trigger to experienced additional force that could interfere with the firing of the device.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot 881a33 number, and no non-conformances were identified.
|