Model Number OER-PRO
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/08/2022
Event Type  malfunction  
Event Description
The customer reported to olympus technical assistance center (tac), a scope was reprocessed with the endoscope reprocessor, after being used with black dye.It was noted, the scope was reprocessed twice and when scope was hung to dry, it was noticed, black dye was coming out of the distal tip.There was no harm or user injury reported due to the event.The scope that was reprocessed with the subject device is addressed in (patient identifier: (b)(6), serial number: (b)(4), model number: cf-hq190l) and a report will be submitted.
 
Manufacturer Narrative
The subject device has not been returned to olympus for evaluation.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and because the device wasn't returned for analysis, the definitive root cause of the reprocessing issue could not be determined.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿always preclean each endoscope immediately after the examination.If precleaning is not executed promptly, debris will solidify and may prevent effective reprocessing.Failure to preclean will leave excessive amounts of debris adhering to the endoscope and may compromise the effectiveness of the reprocessing.It may also result in debris accumulating in the endoscope, preventing the endoscope from working correctly.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16106359
MDR Text Key308617379
Report Number9610595-2023-00237
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received04/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-