Model Number 71992-01 |
Device Problems
Device Alarm System (1012); No Audible Alarm (1019)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 12/28/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The date of event is unknown.The date entered is the date abbott diabetes care became aware of the event.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An alarm issue was reported with the adc device.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the required third-party treatment.Additional information regarding the treatment was not provided.There was no report of death or permanent injury associated with this event.
|
|
Event Description
|
An alarm issue was reported with the adc device.The low and high glucose alarms did not sound and customer was not alerted of changes in glucose level.As a result, the required third-party treatment.Additional information regarding the treatment was not provided.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Search Alerts/Recalls
|