Catalog Number ZVM08040 |
Device Problems
Positioning Failure (1158); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/13/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that during a stent placement procedure via a contralateral approach, the stent allegedly failed to deploy.It was further reported that the sheath allegedly kinked.The procedure was completed using another device.There was no reported patient injury.
|
|
Manufacturer Narrative
|
The catalog number identified has not been cleared in the us but is similar to the e-luminexx vascular stent that are cleared in the us.The pro code and 510 k number for the e-luminexx vascular stent are identified.Manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system was not returned and photos were not available for evaluation which leads to inconclusive results.It was indicated that the catheter kink was found during deployment attempts.Based on the information available, the investigation is closed with inconclusive results for failure to deploy and catheter kink.A definitive root cause could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product, it was found that the instructions for use sufficiently address the potential risks.With regards to general warnings, the instructions for use states that "should unusual resistance be felt at any time during the procedure, the entire system (introducer sheath or guiding catheter and stent delivery system) should be removed as a single unit".Regarding potential damages, the instructions for use states: "visually inspect the bard e-luminexx vascular stent to verify that the device has not been damaged due to shipping or improper storage.Do not use damaged equipment".(expiration date: 12/2023).Device not returned.
|
|
Manufacturer Narrative
|
H10: the file was reassessed for reportability and determined to be no longer reportable.Since an initial mdr was submitted, therefore, the file will remain assessed as a malfunction.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the e-luminexx vascular stent that are cleared in the us.The pro code and 510 k number for the e-luminexx vascular stent are identified in d2 and g4.H10: d4 (expiration date: 12/2023), g3.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that during a stent placement procedure via a contralateral approach, the stent allegedly failed to deploy.It was further reported that the sheath allegedly kinked.The procedure was complete using another device.There was no reported patient injury.
|
|