• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE; DERMATOME PNEUMATICALLY-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. DERMATOME AN HANDPIECE; DERMATOME PNEUMATICALLY-POWERED Back to Search Results
Model Number N/A
Device Problems Component Missing (2306); Material Integrity Problem (2978); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2022
Event Type  malfunction  
Manufacturer Narrative
This incident was recorded under (b)(4).Upon completion of the investigation a final/supplemental report will be submitted.(b)(6).
 
Event Description
It was reported that the device was sent in for preventative maintenance and noted to need repairs for missing screw, defective sleeve bearings.There was no patient involvement as a result no harm or delay were reported.No adverse event was reported as it relates to this event.
 
Event Description
There is no additional information available regarding the event.
 
Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the device was in calibration but had a calibration issue due to a bad bearing and the width plate screw was missing.The device was returned unrepaired as no response was received for the quote.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.There was no event as the device was returned for preventive maintenance only.It is confirmed the device was in need of repair.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.G2 country: austria.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DERMATOME AN HANDPIECE
Type of Device
DERMATOME PNEUMATICALLY-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key16113572
MDR Text Key307094432
Report Number0001526350-2023-00019
Device Sequence Number1
Product Code GFD
UDI-Device Identifier00889024464834
UDI-Public(01)00889024464834(11)190322(10)64367743
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number88710100
Device Lot Number64367743
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/19/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-