Model Number N/A |
Device Problems
Component Missing (2306); Material Integrity Problem (2978); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/19/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This incident was recorded under (b)(4).Upon completion of the investigation a final/supplemental report will be submitted.(b)(6).
|
|
Event Description
|
It was reported that the device was sent in for preventative maintenance and noted to need repairs for missing screw, defective sleeve bearings.There was no patient involvement as a result no harm or delay were reported.No adverse event was reported as it relates to this event.
|
|
Event Description
|
There is no additional information available regarding the event.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).Review of the most recent repair record determined the device was in calibration but had a calibration issue due to a bad bearing and the width plate screw was missing.The device was returned unrepaired as no response was received for the quote.Review of the device history record identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.There was no event as the device was returned for preventive maintenance only.It is confirmed the device was in need of repair.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.G2 country: austria.
|
|