Model Number CD3369-40Q
Device Problems High impedance (1291); Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2022
Event Type  malfunction  
Event Description
It was reported that during implant device interrogation revealed high defibrillation impedance, high pacing impedance due to an issue with connection to the device header on the implantable cardioverter defibrillator (icd).The physician elected to explant and replace the icd.The patient was in stable condition.
 
Manufacturer Narrative
The reported field events of connection issue and high impedance were not reproduced in the lab.Electrical, longevity, and visual analysis was normal with no anomalies found.A device history record (dhr) review was performed and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16117555
MDR Text Key307505456
Report Number2017865-2023-00562
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508377
UDI-Public05414734508377
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCD3369-40Q
Device Catalogue NumberCD3369-40Q
Device Lot NumberA000127347
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/19/2022
Initial Date FDA Received01/06/2023
Supplement Dates Manufacturer Received01/20/2023
Supplement Dates FDA Received02/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
1458Q/75 LEFT VENTRICLE; 2088TC/52 ATRIAL; LDA210Q/58 RIGHT VENTRICLE
Patient Age77 YR
Patient SexFemale
-
-