• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. 36MM LINER IMPACTOR; EXTREMITY, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. 36MM LINER IMPACTOR; EXTREMITY, INSTRUMENT Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/08/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up report will be submitted.
 
Event Description
It was reported that the pin broke off of the impactor when implanting in patient, and the piece was retained by the patient.No additional information was available.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.H6: component codes: mechanical (g04) - impactor.Visual examination of the provided pictures identified the peg on the end of the impactor was broken off.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
36MM LINER IMPACTOR
Type of Device
EXTREMITY, INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16122587
MDR Text Key306920694
Report Number0001822565-2023-00080
Device Sequence Number1
Product Code HSD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00430902900
Device Lot Number62483689
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/12/2022
Initial Date FDA Received01/09/2023
Supplement Dates Manufacturer Received02/21/2023
Supplement Dates FDA Received03/01/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/02/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
-
-