It was reported that while the patient was on the centrimag device in the intensive care unit, the nursing staff noted that when they would move the centrimag flow probe that was hooked to the console, it wouldn't alarm with interrupted flow notification.The flow probe was swapped out for a new one and the console alarmed as expected.The flow probe also was not measuring flows correctly as the new flow probe was 1l/min off from what the original flow probe was reading.
|
Correction: the initial report was incorrectly submitted with a cfn-based manufacturer report (mfr) number: 2916596-2022-16199.The mfr number should have been fei-based with the 10 digit fei being 3003306248.Manufacturer's investigation conclusion: the reported events of flow alarms not activating and the flow probe not measuring the flow readings properly were not confirmed as the reported issues were not reproduced during testing of the returned centrimag flow probe.The returned centrimag flow probe was connected to a known working test centrimag system without any issues.The flow probe sensor was disconnected from the mock circulatory loop multiple times and a flow alarm activated each time, as expected.The flow readings were then compared between the returned flow probe and the test flow probe, and the flow readings appeared to be similar.The returned centrimag flow probe functioned as intended throughout testing.The root cause of the reported events could not be conclusively determined through this analysis.The device history records were reviewed and the records revealed that the centrimag flow probe was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction.".The 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support.".The 2nd generation centrimag system operating manual section 12.1 ¿ "appendix i ¿ primary console alarms and alerts" cover all alarms (auditory and visual), including flow related alarms, and the appropriate actions to take if the issue does not resolve.No further information was provided.The manufacturer is closing the file on this event.
|