Model Number 720250
Device Problem Unintended Deflation (4061)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/15/2022
Event Type  malfunction  
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
The customer reported the probe balloon leaked during use.It was necessary to exchange the probe.
 
Manufacturer Narrative
Additional information: the device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported condition.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.A sample analysis could not be performed because no photo or sample was available for evaluation.Based on the present information, a root cause cannot be determined.Based on the historical review, a corrective and preventative action (capa) was generated to the supplier to further investigate this issue.This complaint will be used for tracking and trending purposes.
 
Manufacturer Narrative
Additional information: h4 device manufacture date was added the device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported condition.One sample was received for the evaluation.A visual inspection and functional testing were performed.The balloon was inflated with water and a leak was detected.A corrective and preventative action (capa) was opened to further investigate this issue.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALLOON SLG 20FR X 2.5CM
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HELATH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16129858
MDR Text Key308114320
Report Number9612030-2023-03516
Device Sequence Number1
Product Code KNT
Combination Product (y/n)N
Reporter Country CodeBR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number720250
Device Catalogue Number720250
Device Lot Number1919725564
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/24/2019
Type of Device Usage A
Patient Sequence Number1
-
-