• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS NORTH AMERICA LLC WISP MASK; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS NORTH AMERICA LLC WISP MASK; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number 1094051
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Failure of Implant (1924)
Event Date 08/04/2023
Event Type  malfunction  
Manufacturer Narrative
The instructions for use will include the following warnings: the mask assembly contains magnets.Some medical devices may be affected by magnetic fields.The magnetic clips in this mask should be kept at least 6 inches (approx.15.24 cm) away from any active medical device with special attention to implanted devices such as pacemakers, defibrillators and cochlear implants.Do not use in or near magnetic resonance imaging (mri) equipment.Keep unassembled magnetic headgear clips out of reach of children.In case of accidental swallowing, seek medical assistance immediately.Mask and headgear still in use, not returning.
 
Event Description
The manufacturer became aware of an allegation that a user's dental implant fell out while using a wisp mask with magnets.The user states about two years ago her dental implant fell out and she saw a specialist to have the tooth removed.There was no serious patient harm or injury reported.No product is returning for investigation.The user was advised to remove the magnets in her mask and to order the plastic clips for the replacement.This is an initial final report.The manufacturer will continue to monitor complaints for similar issues.The manufacturer concludes no further action is needed at this time.
 
Manufacturer Narrative
Corrected data includes: type of reported complaint being changed to product problem.
 
Manufacturer Narrative
On previously submitted report, box b: "adverse event or product problem" was incorrect.In this report, box b: "adverse event or product problem" has been updated/corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WISP MASK
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
PHILIPS NORTH AMERICA LLC
1001 murry ridge lane
ste a
murrysville PA 15668
Manufacturer (Section G)
PHILIPS NORTH AMERICA LLC
1001 murry ridge lane
ste a
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key16131263
MDR Text Key307196297
Report Number2518422-2023-01176
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121631
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1094051
Device Catalogue Number1094051
Was Device Available for Evaluation? No
Date Manufacturer Received10/26/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Removal/Correction NumberRES90705¿¿¿¿¿¿
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age88 YR
Patient SexFemale
-
-