• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR.; LIGATION / VESSEL CLIPS Back to Search Results
Model Number PL579T
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
It was reported that there was an issue with pl579t - challenger ti-p ml-ligat.Clips 12 cartr.According to the complaint description, the end of the challenger cartridge snapped off inside the patient during a robotic prostatectomy.They were however able to retrieve it and no harm was done.An additional medical intervention was necessary.Additional information was not provided.The adverse event is filed under aag reference (b)(4).
 
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Manufacturer Narrative
Additional information: h6 - codes updated.Investigation: an investigation method could not be applied, because products were not provided to us for investigation.In the picture provided in sap, the nose of the product is fractured.Batch history review: the device quality and manufacturing history records (dhr) have been checked for all leading device(s) lot numbers and the products found to be according to our specification valid at the time of production.There are no similar complaints against the same lot number(s) with this error pattern.Explanation, rationale, conclusion and root cause: unfortunately, without the product, and only with the picture archived in sap database, we cannot determine the exact cause.According to the quality standard, a material defect and a production error can most probably be excluded.If further investigations are required, the product should be provided for examination.A capa was already initiated for various failure modes of the challenger system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHALLENGER TI-P ML-LIGAT.CLIPS 12 CARTR.
Type of Device
LIGATION / VESSEL CLIPS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key16134023
MDR Text Key307039209
Report Number9610612-2022-00397
Device Sequence Number1
Product Code FZP
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K081031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPL579T
Device Catalogue NumberPL579T
Device Lot Number52797612
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/30/2022
Initial Date FDA Received01/10/2023
Supplement Dates Manufacturer Received02/06/2023
Supplement Dates FDA Received02/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-