• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NOVO NORDISK A/S, MEDICAL SYSTEMS NOVOPEN ECHO; INSULIN DELIVERY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NOVO NORDISK A/S, MEDICAL SYSTEMS NOVOPEN ECHO; INSULIN DELIVERY DEVICE Back to Search Results
Model Number N/A
Device Problem Device Displays Incorrect Message (2591)
Patient Problem Hyperglycemia (1905)
Event Date 12/27/2022
Event Type  Injury  
Event Description
Event verbatim [preferred term] (related symptoms if any separated by commas) hyperglycaemia [hyperglycaemia].The dial showed 0 after administraion of novorapid at 5iu right before lunch.However, the memory display showed "end".[device information output issue].Case description: this serious spontaneous case from japan was reported by a consumer as "hyperglycaemia(hyperglycaemia)" beginning on (b)(6) 2022, "the dial showed 0 after administraion of novorapid at 5iu right before lunch.However, the memory display showed "end".(device information output issue)" with an unspecified onset date, and concerned a female patient who was treated with novopen echo (insulin delivery device) from unknown start date for "device therapy", , novorapid chu penfill (insulin aspart) solution for injection, 100 iu/ml (dose, frequency & route used - unk, subcutaneous) from unknown start date for "drug use for unknown indication", patient's height, weight and body mass index was not reported.Dosage regimens: novopen echo: novorapid chu penfill: not reported to not reported, 27-dec-2022 to not reported, not reported to not reported; medical history was not provided.Concomitant products included - tresiba chu penfill(insulin degludec) solution for injection, 100 iu/ml, novopen echo(insulin delivery device) n/a on (b)(6) 2022 the dial showed 0 after administraion of novorapid at 5iu right before lunch.However, the memory display showed "end".Patient did not remember such as whether she dropped novopen echo (red color).The date red color of novopen echo was given to her was unknown.Thereafter, there was an unfortunate incident around her.She spent different days from usual and was in stimulated condition.On the same date, her blood sugar level was 539 mg/dl when measuring it a little past 5:00 pm.When she set the units of novorapid at 2iu and pushed the infusion button of novopen echo (blue color) which was used for tresiba, the memory display showed "2".She commented that injection had not been performed successfully due to the error of novopen echo (red color).Batch numbers: novorapid chu penfill and novopen ech was requested.Action taken to novopen echo was not reported.Action taken to novorapid chu penfill was reported as unknown.The outcome for the event "hyperglycaemia(hyperglycaemia)" was unknown.The outcome for the event "the dial showed 0 after administraion of novorapid at 5iu right before lunch.However, the memory display showed "end".(device information output issue)" was not reported.In order to protect the safety of patient it will, in rare cases, be required to disassemble the medical device immediately in a way where it is not subsequently possible to reassemble it (e.G.Destructive testing or altering of the medical device).The disassembled medical device will be stored with the same retention period as other complaint samples.Preliminary manufacturer comment: (b)(6) 2023: the suspected device novopen echo has not been returned to novo nordisk yet to evaluate if it works according to set specification and intended use.With very limited information regarding handling of the suspected device reported, dropping the device in the case, it is not possible to elucidate a clear root cause for the experienced adverse event of hyperglycaemia.Additional dosage regimens: suspect product 2.Dose, frequency & route used 3.Therapy dates (if unknown, give duration) 6.Lot # 7.Exp.Date.#1 novorapid chu penfill regimen # 2 5iu right before lunch., subcutaneous (b)(6) 2022 to unk #1 novorapid chu penfill regimen # 3 2iu, subcutaneous.
 
Event Description
Case description: investigational result: novorapid chu penfill - batch unknown no investigation was possible, because neither sample nor batch number was available.Novopen echo- batch unknown no investigation was possible, because neither sample nor batch number was available.Final manufacturer comment: 02-mar-2023: the suspected device novopen echo has not been returned to novo nordisk for investigation.Batch number of the device unavailable despite repeated efforts to find the same.No batch trend analysis or reference sample analysis performed.No other confounding factors identified.With the available limited information regarding the handling of the suspected device, it is not possible to identify a clear root cause in relation to functionality of novopen echo.It was reported that memory display showed "end".That means, the device battery has been expired.The user can detect that the display does not work as intended; however, the pen can still be used to deliver the needed insulin.It is estimated that the fault has no impact on the patient's health or treatment.H3 continued: evaluation summary novopen echo- batch unknown no investigation was possible, because neither sample nor batch number was available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOVOPEN ECHO
Type of Device
INSULIN DELIVERY DEVICE
Manufacturer (Section D)
NOVO NORDISK A/S, MEDICAL SYSTEMS
hilleroed,
DA 
Manufacturer (Section G)
NOVO NORDISK A/S
krogshoejvej 55
bagsvaerd, 2880
DA   2880
Manufacturer Contact
p.o. box 846
plainsboro, NJ 08536
8007276500
MDR Report Key16134584
MDR Text Key307061447
Report Number9681821-2023-00007
Device Sequence Number1
Product Code FMF
Combination Product (y/n)Y
Reporter Country CodeJA
PMA/PMN Number
K123766
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number185459
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/27/2022
Initial Date FDA Received01/10/2023
Supplement Dates Manufacturer Received02/21/2023
Supplement Dates FDA Received03/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
NOVOPEN ECHO (INSULIN DELIVERY DEVICE); TRESIBA CHU PENFILL (INSULIN DEGLUDEC)
Patient Outcome(s) Other;
Patient SexFemale
-
-