Model Number WA20021A
Device Problems Break (1069); No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2022
Event Type  malfunction  
Manufacturer Narrative
The referenced scope was returned to the olympus service center.The service inspection confirmed the customer¿s reported issue, broken lens inside the optical system attributing to a blurry image.No moisture was observed inside the optical system.The investigation is still in progress.However, if additional information becomes available, this report will be supplemented accordingly.In general, the end-user is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.
 
Event Description
Olympus was informed via repair request that the customer autoclavable telescope has ¿no image, lens cracked/broken inside¿.The customer reported problem was found at inspection before use.No death or injury and no impact to person or other was reported to olympus.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation (h6/h10).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.It is likely the event occurred due to the use of excessive force.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 4 MM, 12°, LONG
Type of Device
AUTOCLAVABLE TELESCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, de 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16136239
MDR Text Key307071358
Report Number9610773-2023-00176
Device Sequence Number1
Product Code GCP
UDI-Device Identifier04042761051422
UDI-Public04042761051422
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K790071
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA20021A
Device Catalogue NumberWA20021A
Device Lot Number576397
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/31/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/05/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-