Model Number WA20021A |
Device Problems
Break (1069); No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The referenced scope was returned to the olympus service center.The service inspection confirmed the customer¿s reported issue, broken lens inside the optical system attributing to a blurry image.No moisture was observed inside the optical system.The investigation is still in progress.However, if additional information becomes available, this report will be supplemented accordingly.In general, the end-user is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.
|
|
Event Description
|
Olympus was informed via repair request that the customer autoclavable telescope has ¿no image, lens cracked/broken inside¿.The customer reported problem was found at inspection before use.No death or injury and no impact to person or other was reported to olympus.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation (h6/h10).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be determined.It is likely the event occurred due to the use of excessive force.Olympus will continue to monitor field performance for this device.
|
|