ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Model Number ER320
Device Problems Mechanical Problem (1384); Failure to Form Staple (2579); No Apparent Adverse Event (3189)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # unk.Additional information was requested and the following was obtained: "the sales rep checked the device on (b)(6).The sales rep fired the device, malformed clip came out.The remaining clip turned 0, and the device locked out." this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that during a laparoscopic mile¿s operation, two clips came out together after 9th firing.Two clips were retrieved, and attempted to fire the device again, but the clip did not close and fell off.The same issue occurred for the next clip as well.In total, four clips would not close properly.The patient was severely obese and has a medical history of cerebral infarction.Slight bleeding occurred, and a blood transfusion was not performed.No adverse consequences to the patient.Another device was used to complete the case.
 
Manufacturer Narrative
(b)(4).Date sent: 2/9/2023.D4: batch # x95m3p.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the er320 device was returned for analysis and upon inspection, the jaws were found to be in a yielded condition.In an attempt to replicate the reported incident, the device was functionally evaluated.Upon firing of the device it was noted to be empty and locked out.The instrument is designed to lock out after all the clips have been fired; therefore a potential cause of the customer reported experience is the firing of all of the clips and the instrument "will not fire" (activation of the lock out mechanism).In order to evaluate the condition of the internal components of the device, it was disassembled.Upon disassembling, no anomalies were found.It is known from the history of the device that the condition of the jaws may lead to dropping/ejected clips.The event reported was confirmed and it is related to improper use of the device.Possible causes for the found condition of the yielded jaws maybe if the device is closed over an existing hard object or clip placing stress on the jaws causing them to distort or yield and not return to their original dimensions/position or excessive application of torque to the jaws when positioning the device on a vessel.Please reference the instruction for use for more information.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch x95m3p number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key16137843
MDR Text Key308627499
Report Number3005075853-2023-00245
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public10705036012580
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberER320
Device Catalogue NumberER320
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2023
Is the Reporter a Health Professional? No
Date Manufacturer Received01/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/19/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-