MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC. SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC. SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 595000-001
Device Problems Appropriate Term/Code Not Available (3191); Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2022
Event Type  malfunction  
Manufacturer Narrative
The freedom driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
 
Event Description
The customer, a syncardia authorized distributor, reported that freedom driver #5291 went into red and irreversible alarm immediately without any reason.
 
Manufacturer Narrative
Alarm history and patient data file review found no new alarms recorded in the freedom driver's data file.Visual inspection of internal and external components revealed no damage or abnormalities.Freedom driver passed all incoming functional testing.Additional functional testing included a power cycle test, where left and right batteries are alternately replaced 10 times each.No alarms were produced during this test.Testing was not able to replicate the condition, no fault alarms occurred during this test.The driver's power adaptor and/or freedom batteries in use at the moment of the reported issue were not returned with the driver therefore; not included in the investigation.Failure investigation for this complaint could not confirm the reported issue via alarm history data review nor functional testing.The complaint was not replicated via functional testing.The root cause of the customer reported alarm was unable to be determined.Failure investigation identified no test failure, damage or abnormalities that could have contributed to the reported alarm.Driver functioned as intended.Patient was switched to back-up driver without any reported adverse impact.This issue will be monitored as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.(if new or additional information is received in the future, syncardia will file a follow-up mdr.) (b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key16137992
MDR Text Key308374304
Report Number3003761017-2022-00154
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 06/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received01/19/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-