The technician was trained in april 2021 on operating the autosection, and worked with the instrument almost daily and is considered very experienced.The customer was instructed to stop usage of the instrument until sakura inspected the instrument.A sakura technical engineer visited the customer on (b)(6) 2023 and conducted a technical and visual inspection, but no abnormalities were found.The sakura application specialist visited the customer on (b)(6) 2023 to test the instrument and concluded that the autosection worked as intended within specification.Preliminary analysis of the autosection log files from the moment of incident do not show any abnormalities with the operation of the autosection.
|
Sakura finetek europe was notified of an incident on (b)(6) 2022, which occurred in (b)(6) on (b)(6) 2022 with the tissue-tek autosection automated microtome instrument, ser#: (b)(4).A (b)(6) customer reported a technician was injured by their autosection ivd device, and asked for technical inspection of the instrument.The customer reported that the technician used the foot pedal to stop the autosection from operation (this is intended behavior).According to the technician, the instrument stopped operation (slicing motions to cut tissue in a cassette) and wanted to remove the cassette, but during this act, the instrument started to move again without user interaction and the incident happened.The fingertip of the technician was severed and placed back with surgery after visiting the emergency room of the hospital.
|