Model Number LS1500 |
Device Problems
Energy Output Problem (1431); Difficult to Open or Close (2921)
|
Patient Problems
Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Date 12/20/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during procedure, the device jaws were "tearing tissue" when activated via energy.There was no regrasp alert, end tone was heard, and there was a sign of energy delivery.Used similar device and worked successfully.The generator settings was normal.
|
|
Manufacturer Narrative
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device produced only a partial seal of the vessel, and the jaws of the device were difficult to open.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during laparoscopically assisted vaginal hysterectomy (lavh), the device¿s jaw was unable to unlock and jaws were "tearing tissue" when activated via energy.There was no regrasp alert, end tone was heard, and there was a sign of energy delivery.The device was applied on 7mm thick uterine artery.The generator settings was normal.A new device was opened and used on the same generator with no issues.Patient had bleeding for approximately 200 ml and no blood transfusion needed.
|
|