BECTON, DICKINSON & CO. (SPARKS) BD TOTALYS SLIDEPREP; PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED Back to Search Results
Model Number 491346
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2022
Event Type  malfunction  
Event Description
It was reported that bd totalys slideprep contamination is occurring.The following information was provided by the initial reporter: contamination is occurring.
 
Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported that bd totalys slideprep contamination is occurring.The following information was provided by the initial reporter: contamination is occurring.
 
Manufacturer Narrative
H.6 investigation summary: complaint reports contamination on slideprep (catalog number 491346) serial number (b)(6).Complaint alleges instrument contamination due to bundle touching slides, no erroneous results were reported.Service adjusted the quad bundle z max height and realigned the quad arm.Root cause attributed to misalignment of quad arm.This complaint is a confirmed failure of the instrument based on the service investigation.Review of device history record for instrument serial number, ppla300112 is not required because this complaint does not allege an early life failure or failure at installation and has changed configuration since release from manufacturing due to service repairs/pms.Device was installed on 09/24/2013.Service history review was performed for the instrument ppla300112, and no additional work orders were observed for the complaint failure mode reported.Review of risk management files confirms there are no new or modified risks associated with this failure mode.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD TOTALYS SLIDEPREP
Type of Device
PROCESSOR, CERVICAL CYTOLOGY SLIDE, AUTOMATED
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer (Section G)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16142695
MDR Text Key308079364
Report Number1119779-2023-00022
Device Sequence Number1
Product Code MKQ
UDI-Device Identifier00382904913465
UDI-Public00382904913465
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P970018
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number491346
Device Catalogue Number491346
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-