• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. ENSEAL X1 CURVED 37CM SHAFT; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. ENSEAL X1 CURVED 37CM SHAFT; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number NSLX137C
Device Problems Structural Problem (2506); Material Integrity Problem (2978); Firing Problem (4011)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch #: x94l7c.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with the upper jaw broken at the closure slot.Not all pieces were returned with the device.The device was connected to the generator and it was recognized.Because the jaw was damaged not all functional testing could be performed with the generator.The jaws could not close fully.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event.A manufacturing record evaluation was performed for the finished device batch x94l7c, and no non-conformances related to the reported complaint condition were identified.
 
Event Description
It was reported that, during total laparoscopic hysterectomy, there was a difficulty in cutting from the beginning.Handle became not to open, and it was found that the jaws were damaged on the mayo table.No pieces fell into the patient.Gen11 was used.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSEAL X1 CURVED 37CM SHAFT
Type of Device
ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO SURGERY, INC. (CINCINNATI)
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key16145440
MDR Text Key308609553
Report Number3005075853-2023-00219
Device Sequence Number1
Product Code GEI
UDI-Device Identifier10705036015468
UDI-Public10705036015468
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K172580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNSLX137C
Device Catalogue NumberNSLX137C
Device Lot NumberX94L7C
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2022
Date Manufacturer Received12/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/22/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Treatment
GENERATOR.
-
-