ETHICON ENDO-SURGERY, LLC. ENSEAL X1 CURVED 37CM SHAFT; ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Model Number NSLX137C |
Device Problems
Structural Problem (2506); Material Integrity Problem (2978); Firing Problem (4011)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/28/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Batch #: x94l7c.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the device was received with the upper jaw broken at the closure slot.Not all pieces were returned with the device.The device was connected to the generator and it was recognized.Because the jaw was damaged not all functional testing could be performed with the generator.The jaws could not close fully.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event.A manufacturing record evaluation was performed for the finished device batch x94l7c, and no non-conformances related to the reported complaint condition were identified.
|
|
Event Description
|
It was reported that, during total laparoscopic hysterectomy, there was a difficulty in cutting from the beginning.Handle became not to open, and it was found that the jaws were damaged on the mayo table.No pieces fell into the patient.Gen11 was used.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
|