• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENUSA DE MEXICO S.A. DE C.V. TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENUSA DE MEXICO S.A. DE C.V. TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number PN-004 075
Device Problems Fluid/Blood Leak (1250); Electrical Shorting (2926)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2022
Event Type  malfunction  
Manufacturer Narrative
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.Electrodes 1-4 met specifications during electrical testing with no open circuits detected.However, multiple short circuits were detected, consistent with the reported signal issue and impedance issue.Fluid was noted outside of the irrigation tubing but within the tygon tubing proximal to the catheter handle, consistent with the reported leak.In addition, dried saline was noted within the connector plug, consistent with the short circuits detected.Further testing revealed a leak at the luer/irrigation tubing junction due to a gap in the adhesive between the irrigation tube and the proximal tube inside the luer lock, allowing fluid to flow outside of the irrigation tubing and into the proximal tubing.This is consistent with the dried saline noted within the connector plug.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The gap in the adhesive between the irrigation tube and the proximal tube inside the luer lock was determined to be supplier related.
 
Event Description
This report is to advise of an event observed during analysis which revealed multiple short circuits and a leak.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX  32470
Manufacturer (Section G)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX   32470
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16156946
MDR Text Key307281249
Report Number9680001-2023-00003
Device Sequence Number1
Product Code OAE
UDI-Device Identifier07640157990040
UDI-Public07640157990040
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/25/2023
Device Model NumberPN-004 075
Device Catalogue NumberPN-004075
Device Lot Number7852127
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/22/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-