Catalog Number 3095040 |
Device Problem
Difficult to Open or Remove Packaging Material (2922)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/13/2022 |
Event Type
malfunction
|
Event Description
|
The packaging of the outer layer was stuck with inner layer, nurse was unable to peel open the packaging.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Dmf# - 13704.Trade name gentamicin sulphate.Active ingredient(s) gentamicin sulphate.Dosage form - powder.Strength 1.0g active in our cements.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary : the device associated with this complaint was not received for examination.However we have found that the cement powder inner bag is overlapping with the paper outer foil for the same product in different batches/lot#.Additionally we have not found any compromise in the sterility and/or the product.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Device history lot:a search of the depuy nonconformance (nc) quality system found no past nc¿s associated with the current complaint issue.However an nc has been raised to address the current complaint condition.
|
|