• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY CMW - 9610921 SMARTSET GHV GENTAMICIN 40G; BONE CEMENT : BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY CMW - 9610921 SMARTSET GHV GENTAMICIN 40G; BONE CEMENT : BONE CEMENT Back to Search Results
Catalog Number 3095040
Device Problem Difficult to Open or Remove Packaging Material (2922)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2022
Event Type  malfunction  
Event Description
The packaging of the outer layer was stuck with inner layer, nurse was unable to peel open the packaging.
 
Manufacturer Narrative
Product complaint # (b)(4).Dmf# - 13704.Trade name gentamicin sulphate.Active ingredient(s) gentamicin sulphate.Dosage form - powder.Strength 1.0g active in our cements.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint (b)(4).Investigation summary : the device associated with this complaint was not received for examination.However we have found that the cement powder inner bag is overlapping with the paper outer foil for the same product in different batches/lot#.Additionally we have not found any compromise in the sterility and/or the product.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Device history lot:a search of the depuy nonconformance (nc) quality system found no past nc¿s associated with the current complaint issue.However an nc has been raised to address the current complaint condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSET GHV GENTAMICIN 40G
Type of Device
BONE CEMENT : BONE CEMENT
Manufacturer (Section D)
DEPUY CMW - 9610921
cornford rd.
blackpool IN FY4 4 QQ
UK  FY4 4QQ
Manufacturer (Section G)
DEPUY CMW - 9610921
cornford rd
blackpool FY4 4 QQ
UK   FY4 4QQ
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16157133
MDR Text Key308528649
Report Number1818910-2023-01156
Device Sequence Number1
Product Code MBB
Combination Product (y/n)Y
Reporter Country CodeSN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3095040
Device Lot Number3750269
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/12/2022
Initial Date FDA Received01/13/2023
Supplement Dates Manufacturer Received01/17/2023
02/03/2023
Supplement Dates FDA Received01/18/2023
02/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-