Model Number CI-1601-04
Device Problems Mechanical Problem (1384); Impedance Problem (2950)
Patient Problems Fall (1848); Head Injury (1879); Hearing Impairment (1881)
Event Date 03/17/2023
Event Type  Injury  
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Event Description
Following a head trauma event the recipient experienced loss of sound perception.External equipment was exchanged, however the issue did not resolve.Lock was achieved during device testing however, impedance measurements suggested open electrodes across the array.Further testing will be conducted to evaluate device position.Advanced bionics is still in the process of obtaining additional information.When additional information is received, a supplemental report will be submitted.
 
Manufacturer Narrative
Imaging revealed proper device placement.Revision surgery will be scheduled.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed and silicone damage was observed on the top cover of the device.This is believed to have occurred during revision surgery.Photographic imaging inspection revealed broken electrode wires within the fantail.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.The scanning electron microscopy (sem) analysis revealed tensile breaks within the fantail.The device passed the residual gas analysis (rga) test.Sem analysis revealed broken electrode wires near the fantail, induced by the trauma reported.It is believed that these breaks caused the open impedances measured at the clinic.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
heidi white
28515 westinghouse place
valencia, CA 91355
MDR Report Key16163627
MDR Text Key307459332
Report Number3006556115-2023-00080
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016878039
UDI-Public(01)07630016878039(11)200113(17)230131
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 01/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date01/31/2023
Device Model NumberCI-1601-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/13/2023
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Not provided
Not provided
Supplement Dates FDA Received02/23/2023
04/24/2023
08/17/2023
10/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age5 YR
Patient SexFemale
-
-