Model Number CI-1601-04 |
Device Problems
Mechanical Problem (1384); Impedance Problem (2950)
|
Patient Problems
Fall (1848); Head Injury (1879); Hearing Impairment (1881)
|
Event Date 03/17/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Event Description
|
Following a head trauma event the recipient experienced loss of sound perception.External equipment was exchanged, however the issue did not resolve.Lock was achieved during device testing however, impedance measurements suggested open electrodes across the array.Further testing will be conducted to evaluate device position.Advanced bionics is still in the process of obtaining additional information.When additional information is received, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Imaging revealed proper device placement.Revision surgery will be scheduled.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the electrode was severed and silicone damage was observed on the top cover of the device.This is believed to have occurred during revision surgery.Photographic imaging inspection revealed broken electrode wires within the fantail.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.The scanning electron microscopy (sem) analysis revealed tensile breaks within the fantail.The device passed the residual gas analysis (rga) test.Sem analysis revealed broken electrode wires near the fantail, induced by the trauma reported.It is believed that these breaks caused the open impedances measured at the clinic.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|