OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°, STERILE, SINGLE USE; HR-RESECTION ELECTRODES
|
Back to Search Results |
|
Model Number WA22306D |
Device Problems
Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/12/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The suspect medical device has not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the patient's outcome and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical device is returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
|
|
Event Description
|
Olympus was informed that during a therapeutic hysteroscopy procedure, the loop wire at the distal end of the hf resection electrode broke off and a fragment fell into the patient.The broken fragment was retrieved using a diagnostic olympus scope and grasping forceps.The intended procedure was then completed but was prolonged by approx.1 hour due to the failure.
|
|
Manufacturer Narrative
|
Additional information: e2 - health professional; e3 - occupation; g2 - report source
device evaluation: the suspect medical device was not returned to olympus for evaluation/investigation since it was reportedly discarded by the user facility.Therefore, the exact cause of the patient¿s outcome and the reported phenomenon could not be conclusively determined and is being judged as unknown.However, the reported failure is a known phenomenon usually caused by use-related wear and tear.The loop wire at the distal end of the hf electrode may wear out during use and may break, burn or melt.Thus, this incident can most likely be attributed to user error.A manufacturing and quality control review could not be performed since basic data of article identification (lot number) are missing.The case will be closed on olympus side with no further actions.However, the reported event/incident will be recorded for trending and surveillance purposes and the user will be informed about the investigation results.
|
|