• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS LUCEA 40; DEVICE, MEDICAL EXAMINATION, AC POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS LUCEA 40; DEVICE, MEDICAL EXAMINATION, AC POWERED Back to Search Results
Model Number ARD568601997
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2023
Event Type  malfunction  
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.Device not returned to manufacturer.
 
Event Description
On 12th january, 2023 getinge became aware of an issue with one of examination lights - lucea 40.It was stated the headlight cover was cracked with missing particles in the attachment the screws to the handle.The screws were no longer holding the device together.We decided to report the issue in abundance of caution as any particles or parts falling off during examination procedure may cause contamination.
 
Manufacturer Narrative
Defective parts transparent plastic cover (ard368606555) and handle interface with fork (ard368605998) were replaced and device back to usage.Based on the information collected, it was established that when the event occurred, the examination light did not meet its specification, since there are cracks found in the light assembly¿s covers, resulting in missing plastic particles ¿ which can be considered as technical deficiency, and in this way the device contributed to the event.The provided information does not indicate if upon the event occurrence, the device was or was not being used for patient treatment.When reviewing reportable events for this type of issue we were able to establish that the received incidents are occurring at a moderate ratio.We have been able to confirm that the investigated issue has never led to serious injury or worse, to our knowledge.According to the analysis performed by the subject matter expert, the probable causes of the breakage are the incompatibility of the cleaning products or abnormal use.Maquet sas recommends visual inspection before use according to the user manual (ifu 01701 rev.12, page 22-23).A daily inspection performed by the user (chipped paint, impact marks and any other damage) will help preventing such event.Maquet sas also recommends to inform the customers about the hazards in cases of non-compliance of these instructions.We believe the related devices are performing correctly in the market.We also believe that if the manufacturer¿s recommendation had been followed the incident could have been avoided.Getinge shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Event Description
Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCEA 40
Type of Device
DEVICE, MEDICAL EXAMINATION, AC POWERED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key16193387
MDR Text Key307570276
Report Number9710055-2023-00035
Device Sequence Number1
Product Code KZF
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARD568601997
Device Catalogue NumberARD568601997
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2023
Initial Date FDA Received01/18/2023
Supplement Dates Manufacturer Received10/18/2023
Supplement Dates FDA Received10/18/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/15/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-