• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG SENTUS PROMRI OTW QP L-85/49; LV LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG SENTUS PROMRI OTW QP L-85/49; LV LEAD Back to Search Results
Model Number 408719
Device Problems Failure to Capture (1081); High impedance (1291); Ambient Noise Problem (2877)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2023
Event Type  malfunction  
Event Description
Homemonitoring reported pacing impedance above 3000 ohm on (b)(6) 2023 with periods of return to baseline.Pacing impedance previously spiked to above 3000 ohm on (b)(6) 2022.In-office follow up testing found all values within range.Noise was observed on the lv channel during postural testing with patient bending forward as if to tie their shoes.Loss of capture was also observed during follow-up.Data to be presented to physician for next steps.Lead currently remains implanted.
 
Manufacturer Narrative
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTUS PROMRI OTW QP L-85/49
Type of Device
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key16205019
MDR Text Key307700062
Report Number1028232-2023-00246
Device Sequence Number1
Product Code OJX
UDI-Device Identifier04035479148959
UDI-Public(01)04035479148959(17)210430
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P070008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Model Number408719
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/17/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexMale
-
-