SMITHS MEDICAL ASD, INC. BIVONA TRACHEOSTOMY TUBES; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
|
Back to Search Results |
|
Catalog Number 67PFSS35 |
Device Problem
Inflation Problem (1310)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).Udi is unknown, 510k is unknown.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.The sample successfully passed functional test.Root cause could not be determined since the sample successfully passed functional test.No corrective actions were taken since the complaint was not confirmed.
|
|
Event Description
|
It was reported that during pre-test, the cuff did not inflate.No patient injury was reported.
|
|