MAUDE Adverse Event Report: TORNIER INC PERFORM REV INSERT SZ 3/4, 39MM DIA, +0; POLYETHYLENE REVERSE SHOULDER PROSTHESIS CUP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC PERFORM REV INSERT SZ 3/4, 39MM DIA, +0; POLYETHYLENE REVERSE SHOULDER PROSTHESIS CUP Back to Search Results
Model Number DWP2390
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Muscle Weakness (1967); Nerve Damage (1979); Paralysis (1997); Loss of Range of Motion (2032)
Event Date 12/20/2022
Event Type  Injury  
Event Description
It was reported the patient was seen on for standard of care follow-up visit.The treating surgeon did not feel a strong contraction in the axillary nerve.Patient could not hold arm in abduction against gravity.Deltoid contraction was minimal.Patient will complete normal physical therapy plan.
 
Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.
 
Manufacturer Narrative
The reported event could be not be confirmed.The device remains implanted and no other evidences were provided that could confirm the allegation.The opinion of the medical expert was requested and stated as following: ¿this axillary nerve dysfunction, that leads to a deltoid muscle weakness is indeed an ae that is not device related yet related to the surgery.Usually it¿s neuropraxia (traction injury without loss of structural nerve continuity) and will resolve spontaneously, but sometimes it doesn¿t, or it doesn¿t completely when the nerve is damaged more severely.It will become clear during follow up.Sometimes electrophysiological investigation, assessing the conduction velocity of the nerve(s) at 6 weeks after the problem started, may provide a recovery prognosis.¿ a device inspection was not possible since the affected device was not returned as it remains implanted in the patient.A review of the labeling and the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.More information as well as the affected device must be available in order to determine the exact root cause of the alleged failure.If any additional information is provided, the investigation will be reassessed.
 
Event Description
It was reported the patient was seen on for standard of care follow-up visit.The treating surgeon did not feel a strong contraction in the axillary nerve.Patient could not hold arm in abduction against gravity.Deltoid contraction was minimal.Patient will complete normal physical therapy plan.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORM REV INSERT SZ 3/4, 39MM DIA, +0
Type of Device
POLYETHYLENE REVERSE SHOULDER PROSTHESIS CUP
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
WRIGHT MEDICAL CORK (TORNIER ORTHOPEDICS IRELAND LTD)
harnetts cross
macroom, co. cork NA
EI   NA
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16210841
MDR Text Key307766694
Report Number0001649390-2023-00017
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00846832084487
UDI-Public00846832084487
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201315
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberDWP2390
Device Catalogue NumberDWP2390
Device Lot NumberAE6265
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/22/2022
Initial Date FDA Received01/19/2023
Supplement Dates Manufacturer Received03/23/2023
Supplement Dates FDA Received04/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/29/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-