• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT SBP DR INDUCTIVE PACEMAKER, JPD; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT SBP DR INDUCTIVE PACEMAKER, JPD; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2152
Device Problems Failure to Interrogate (1332); No Pacing (3268)
Patient Problem Fatigue (1849)
Event Date 12/28/2022
Event Type  Injury  
Event Description
It was reported that the patient experienced fatigue due to loss of pacing observed on the device.The device could not be interrogated.The device was explanted and replaced to resolve the event.The patient was stable following the procedure and there were no adverse consequences.
 
Manufacturer Narrative
The reported events of inability to pace and inability to interrogate were confirmed.The device was received with no telemetry communication and no output.Visual inspection of the header attachment area detected an anomaly between the pre-molded header and titanium case.The device was cut open to enable further testing and the battery was found depleted.Feed through leak test was performed, indicating a device hermeticity breach.This is consistent with feed through damage as a result of fluid intrusion between the header and case, and subsequent fluid ingress to internal electronics.Hybrid circuitry was tested by connecting to an external power source.Test results indicated elevated current drain, consistent with moisture damage, depleting the battery prematurely and resulting in the reported events.A manufacturing process anomaly may have occurred, which resulted in the header bonding anomaly.The device is included in the zenex, assurity, endurity laser adhesion preparation advisory issued by abbott on 20 july 2022 for a subset of devices distributed and implanted outside of the united states.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SBP DR INDUCTIVE PACEMAKER, JPD
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16211794
MDR Text Key307765082
Report Number2017865-2023-02652
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2021
Device Model NumberPM2152
Device Lot NumberP000092032
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/28/2022
Initial Date FDA Received01/19/2023
Supplement Dates Manufacturer Received03/22/2023
Supplement Dates FDA Received04/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/05/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-