Model Number ONYXNG27530UX |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Perforation (2001); Insufficient Information (4580)
|
Event Date 01/06/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
An attempt was made to use one onyx frontier coronary drug eluting stent to treat a mildly calcified lesion in the mid left anterior descending (lad) artery.The device was inspected with no issues noted.Negative prep was performed with no issues noted.The lesion was not pre-dilated.The device did not pass through a previously deployed stent.The stent was deployed in the mildly calcified lesion (no intravascular ultrasound [ivus] was used).It was reported that the stent was deployed at 16 atm for 42 seconds, and then perforation followed.The patient was sent to an outside hospital for covered stent placement.No further injury was reported.
|
|
Manufacturer Narrative
|
Additional information: the lesion was not tortuous with 90% stenosis.There was no resistance when advancing the device to the lesion, and there was no excessive force used.It was thought that potentially the balloon went up faster/grew bigger on the onyx frontier balloon versus the resolute onyx balloon.The patient is alive.Annex d code.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|