• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OTICON MEDICAL - NEURELEC NEURO ZTI COCHLEAR IMPLANT; NEURO ZTI EVO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OTICON MEDICAL - NEURELEC NEURO ZTI COCHLEAR IMPLANT; NEURO ZTI EVO Back to Search Results
Model Number NEURO ZTI EVO
Device Problems Failure to Read Input Signal (1581); Communication or Transmission Problem (2896)
Patient Problem Failure of Implant (1924)
Event Date 12/01/2022
Event Type  malfunction  
Event Description
The decision to explant the patient was made by the medical team.The expected explant date is not known at this stage.In vivo analysis was performed.Oticon medical representative reported the impossibility to connect with the cochear implant despite sound processor, antennas, magnets changes.Electrical artefacts measurements and impedance measurements could not be performed due to the inability of connection with the device.
 
Manufacturer Narrative
While this device has gained initial pma approval, it has not been commercialized, imported or sold in the us and no implantations have been performed as of the date of this report.The implant neuro zti evo was not explanted as of the date of this report.Currently, this event is not a serious injury.Follow-up report will be submitted once the device explanted.The subject device is part of the voluntary field corrective action initiated for neuro zti on october 14th 2021 (international recall #211014).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURO ZTI COCHLEAR IMPLANT
Type of Device
NEURO ZTI EVO
Manufacturer (Section D)
OTICON MEDICAL - NEURELEC
2720, chemin saint-bernard
porte 14
vallauris, 06220
FR  06220
Manufacturer Contact
adrian ternisien
2720, chemin de saint-bernard
porte 14
vallauris, 06220
FR   06220
MDR Report Key16215259
MDR Text Key308000598
Report Number3016743107-2023-00003
Device Sequence Number1
Product Code MCM
UDI-Device Identifier03663227801859
UDI-Public(01)03663227801859(11)201201(17)231201
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
P200021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/01/2023
Device Model NumberNEURO ZTI EVO
Device Catalogue NumberM80185
Device Lot Number20-05396
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-