• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC. SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC. SYNCARDIA FREEDOM DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Catalog Number 595000-001
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2023
Event Type  malfunction  
Event Description
The customer, a syncardia service technician, reported the driver failed the left peak pressure test during incoming.
 
Manufacturer Narrative
The freedom drier will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
 
Manufacturer Narrative
Alarm history and patient data file review found one new alarm recorded in the driver's eeprom: "2d, secondary motor voltage too high".This alarm was produced at syncardia during eeprom extraction.No alarms were recorded related to the current complaint.Visual inspection of internal components revealed no abnormalities.Visual inspection of external components revealed a broken display cover and broken fan cover.Freedom driver failed incoming functional testing for left peak pressure test.Failure investigation for this complaint confirmed the reported issue via patient data file.The complaint could not be replicated via functional/ observational testing.The root cause of the customer reported complaint could not be conclusively determined.No device malfunction nor component failure was identified.Failure investigation identified no test failure, damage, or abnormalities that could have contributed to the customer reported issue.Driver functioned as designed.No patient involvement is related to this routine service related complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.(if new or additional information is received in the future, syncardia will file a follow-up mdr.) (b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer (Section G)
SYNCARDIA SYSTEMS, LLC.
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
aaron meier
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key16224410
MDR Text Key307947023
Report Number3003761017-2023-00001
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 01/03/2023
Initial Date FDA Received01/23/2023
Supplement Dates Manufacturer Received01/03/2023
Supplement Dates FDA Received07/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-