Alarm history and patient data file review found one new alarm recorded in the driver's eeprom: "2d, secondary motor voltage too high".This alarm was produced at syncardia during eeprom extraction.No alarms were recorded related to the current complaint.Visual inspection of internal components revealed no abnormalities.Visual inspection of external components revealed a broken display cover and broken fan cover.Freedom driver failed incoming functional testing for left peak pressure test.Failure investigation for this complaint confirmed the reported issue via patient data file.The complaint could not be replicated via functional/ observational testing.The root cause of the customer reported complaint could not be conclusively determined.No device malfunction nor component failure was identified.Failure investigation identified no test failure, damage, or abnormalities that could have contributed to the customer reported issue.Driver functioned as designed.No patient involvement is related to this routine service related complaint.This issue will be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.(if new or additional information is received in the future, syncardia will file a follow-up mdr.) (b)(4) follow-up report 1.
|