MAUDE Adverse Event Report: APOLLO ENDOSURGERY, INC ORBERA¿ INTRAGASTRIC BALLOON SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APOLLO ENDOSURGERY, INC ORBERA¿ INTRAGASTRIC BALLOON SYSTEM Back to Search Results
Model Number B-4800
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Failure of Implant (1924); Pain (1994); Vomiting (2144); Constipation (3274)
Event Type  malfunction  
Event Description
Patient went to er with the following symptoms: pain, nausea, vomiting, intolerance and constipation.With an endoscopicwhere a needle was used to deflate the balloon and the balloon was removed successfully.
 
Manufacturer Narrative
Initial medwatch submitted to the fda on 23/jan/2023.A review of the device labeling notes the following: the current orbera¿ intragastric balloon system directions for use (dfu) addresses the known and anticipated potential events of "deflation, vomiting and early removal "as follows: "the physiological response of the patient to the presence of the orbera¿ system balloon may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response." "each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms." "complications: possible complications of the use of the orbera¿ system include: gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon.Patients reporting loss of satiety, increased hunger and/or weight gain should be examined endoscopically as this is indicative of an igb deflation.A patient who's deflated (i.E.Collapsed) igb has moved into the intestines must be monitored closely for an appropriate period of time (at least 2 weeks) to confirm its uneventful passage through the intestine.Abdominal or back pain, either steady or cyclic.Deflation and subsequent replacement." deflated device should be removed promptly.Possible adverse events: intestinal obstruction by the igb.An insufficiently filled igb or a leaking igb that has lost sufficient volume may be able to pass from the stomach into the small bowel.It may pass all the way into the colon and be passed with stool.However, if there is a narrow area in the bowel or adhesion formation, which may occur after previous surgery on the bowel, the igb may not pass and could cause a bowel obstruction.If this occurs, surgery or endoscopic removal could be required.Insufficient or no weight loss.Igb deflation (i.E.Collapse) and subsequent replacement.Additional information: the device has not been returned for analysis and attempts to gather more information from the reporter is not visible as this is a literature review, and the author does not have the information for the device.The investigator determined a device history record (dhr) review is not possible.Device evaluation summary: the devicei s not returning ,was discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBERA¿ INTRAGASTRIC BALLOON SYSTEM
Type of Device
INTRAGASTRIC BALLOON SYSTEM
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC
1120 s. captail of texas hwy
bldg 1, ste 300
austin TX 78746
Manufacturer (Section G)
APOLLO ENDOSURGERY COSTA RICA, SRL
coyol free zone
building b 13.3
alajuela, cs CRI
CS   CRI
Manufacturer Contact
adriana russell
1120 s. captail of texas hwy
bldg 1, ste 300
austin, TX 78746
5128523757
MDR Report Key16225994
MDR Text Key308786662
Report Number3006722112-2023-00012
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Literature,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberB-4800
Device Catalogue NumberB-4800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/14/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age24 YR
Patient SexFemale
-
-