Product event summary: the electrophysiology (ep) catheter with lot 16100 was returned and analyzed.During visual inspection of the ep catheter, a breach was observed on the shaft and a kink was observed on the shaft.External visual inspection of the shaft segment, a kink was observed on the shaft at the connection between the shaft and the strain-relief, and a cut on the shaft was observed "from the catheter tip." the catheter smart chip data was reviewed.Data indicated that the catheter was used for nine applications on the reported event date.During system performance testing with the test console, the console terminated the application and generated system notice 50030 (indicating that the safety system detected a high level of refrigerant flow and stopped the injection.) in conclusion, the electrophysiology (ep) catheter failed the returned product inspection due to a breach and a kink that was observed on the shaft.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|