• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE); PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE); PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE) Back to Search Results
Model Number VPR-GW-FLEX14
Device Problem Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/28/2022
Event Type  Injury  
Event Description
Treatments in the heavily calcified superficial femoral artery were performed on low speed and high speed.During removal, the tip of the viperwire advance guide wire became fractured.The fractured tip was successfully snared.The patient was stable.
 
Manufacturer Narrative
Analysis of the reported device is in progress.A supplemental report will be submitted when the device analysis is completed.Csi id: (b)(4).
 
Manufacturer Narrative
The guide wire was returned to csi with a fracture within the spring tip.Scanning electron microscopic analysis revealed evidence of a ductile torsional failure.The reported event was confirmed, however, the root cause was unable to be determined.The material inspection report for this guidewire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: 12524.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE)
Type of Device
PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE)
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
tonia moskalets
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key16241131
MDR Text Key308097369
Report Number3004742232-2023-00021
Device Sequence Number1
Product Code MCW
UDI-Device Identifier10852528005718
UDI-Public(01)10852528005718(17)240831(10)449199-1
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190634
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVPR-GW-FLEX14
Device Catalogue Number7-10041-03
Device Lot Number449199-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2022
Initial Date FDA Received01/25/2023
Supplement Dates Manufacturer Received02/03/2023
Supplement Dates FDA Received02/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-