• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Model Number 367365
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2022
Event Type  malfunction  
Manufacturer Narrative
There were multiple medical device types reported to be involved.The information for each 510(k) number is as follows: common device name: intravascular administration set.Medical device type: fpa.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported when using the bd vacutainer® ultratouch¿ push button blood collection set there was back flow of blood from device to patient.The following information was provided by the initial reporter.The customer stated: there was "flashback and no blood obtained.".
 
Manufacturer Narrative
The following fields have been updated with additional information: d.10 device available for eval? yes.D.10 returned to manufacturer on: 22-feb-2023.H.6.Investigation summary: bd received 26 samples from the customer in support of this complaint.The 26 customer samples, along with 10 retention samples form the bd inventory, were subjected to functional testing to assess functionality of the device.All samples passed testing exhibiting obtainable flash during use.Therefore, this complaint cannot be confirmed based on customer and retain sample testing results.Bd is unable to confirm the customer¿s reported failure mode of flash based on customer sample testing analysis.Based on a review of batch records, no root cause from manufacturing was identified as a contributor.The device history record was reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported when using the bd vacutainer® ultratouch¿ push button blood collection set there was back flow of blood from device to patient.The following information was provided by the initial reporter.The customer stated: there was "flashback and no blood obtained.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® ULTRATOUCH¿ PUSH BUTTON BLOOD COLLECTION SET
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key16241989
MDR Text Key308841497
Report Number1024879-2023-00035
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903673655
UDI-Public50382903673655
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number367365
Device Catalogue Number367365
Device Lot Number2293202
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/03/2023
Initial Date FDA Received01/25/2023
Supplement Dates Manufacturer Received02/24/2023
Supplement Dates FDA Received03/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-