Model Number S-55-150-120-P6 |
Device Problem
Material Separation (1562)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 01/05/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The device has been received; however, investigation has not been completed.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that the procedure was to treat an occluded superficial femoral artery (sfa).The vessel was prepared with a 5.0mm unspecified balloon.The 5.5x150mm supera self-expanding stent tip separated in the anatomy.The tip was attempted to be snared; however, it was unsuccessful.The tip remain inert in the sfa.It was noted that the delivery system was removed under fluoroscopy.There was no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was returned for analysis.The reported material separation was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.As there was no damage noted to the device during the inspection prior to use, the investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the anatomy/other devise and/or inadvertent mishandling resulted in the reported tip separation/noted inner member and tip jacket separations.The treatment(s) appears to be related to the operational context of the procedure as the tip was attempted to be snared.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|
Search Alerts/Recalls
|