Model Number 1011-9050-000
Device Problem No Flow (2991)
Patient Problem Cardiac Arrest (1762)
Event Date 01/03/2023
Event Type  Injury  
Manufacturer Narrative
Ge healthcare investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.No patient information provided to date.Unique identifier: (b)(4).Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.Device evaluation anticipated, but not yet begun.
 
Event Description
The hospital reported a patient connected to an aisys cs2 when mechanical ventilation reportedly stopped.The patient coded when ventilation was performed with an ambu bag while being switched to another anesthesia machine.The clinician resuscitated the patient and was able to complete the case with no further consequence.Ge healthcare will submit a follow-up report when the investigation has been completed.
 
Manufacturer Narrative
Ge healthcare engineering investigated the reported event.The bag/vent switch is designed such that mechanical ventilation is selected if the electrical circuit is open.The device logs support that the bag/vent switch failed such that the microswitch became stuck closed when transitioning back from bag mode and the user was unable to re-initiate mechanical ventilation.The root cause of the loss of mechanical ventilation is the inability for the bag/vent switch to change states from closed to open when transitioning from manual ventilation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AISYS CS2
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key16253148
MDR Text Key308252970
Report Number2112667-2023-00389
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K170872
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1011-9050-000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/09/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/28/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-