EDWARDS LIFESCIENCES EDWARDS CERTITUDE DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
|
Back to Search Results |
|
Model Number 9630TA26 |
Device Problems
Burst Container or Vessel (1074); Detachment of Device or Device Component (2907)
|
Patient Problem
Unspecified Tissue Injury (4559)
|
Event Date 01/09/2023 |
Event Type
Death
|
Manufacturer Narrative
|
Investigation is ongoing.
|
|
Event Description
|
As reported by our affiliates in germany, it was a case of an implant of a 26mm sapien 3 ultra valve in aortic position by transapical approach.Previous balloon aortic valvuloplasty was not performed.During the valve implantation, no extra volume was added to the balloon prior to inflation and 23 ml with the slow inflation technique was used.After the balloon was fully inflated, it burst, and the distal tip was separated.Despite this, the valve was deployed, and the position was acceptable.The proximal part of the balloon and the handpiece could be withdrawn in the introducer sheath.The introducer sheath was removed as a single unit, but the distal part of the balloon remained in the ascending aorta with the guidewire.The remaining balloon portion was fixed by a coil on the wire and the coil wire was wrapped around the guidewire.Due to the porcelain aorta and the calcified iliac access, the recovery had to be done by transapical access.A check flo extra large 24f sheath cook was used without dilator to expand the apex insertion and recover the distal part of the balloon.The check flo sheath and the distal balloon part were removed as a single unit.Unfortunately, closure of the apex was not possible due to tearing of the tissue and loss of sutures with pledgets after the removal of the cook check flo sheath.The patient passed away.As per medical opinion, the reason for the balloon burst was high degree of annulus and leaflets calcification and porcelain aorta.
|
|
Manufacturer Narrative
|
Updated h6-type of investigation, investigation findings, investigation conclusions.The returned device was evaluated, and the following was observed: radial balloon burst was confirmed at the distal portion.There was no missing balloon material.The inflation balloon single wall thickness was measured along the edges of the burst location.All measurements taken of the balloon single wall thickness met the specification.The complaints for balloon burst and distal tip component separation during use were confirmed by visual inspection of the returned device and imagery pro vided.However, no manufacturing non-conformance was identified during the evaluation.Dimensional inspection of the returned balloon revealed that the balloon wall thickness was within specification.No visual abnormalities were observed on the returned sample.A review of edwards lifesciences risk management documentation was performed for this case.The reported events are an anticipated risk of the transcatheter heart valve (thv) procedure, additional assessment of the failure modes is not required at this time.An existing technical summary has been documented for root cause analysis on balloon bursts in a calcified landing zone.The technical summary provides a rationale as to why it is unlikely that a product defect or manufacturing non-conformance contributed to this type of event, including factors on why deployment of balloons on transcatheter heart valve delivery systems are subject to increased risk of burst in a calcified landing zone.As per 3mensio report provided, high level of calcification was observed.The presence of calcification can create a challenging anatomy for balloon inflation.While the balloons are sufficiently designed and tested to ensure the burst pressure is at or above the rated burst pressure, calcified nodules can compromise the structure of the balloon wall via following mechanisms such as puncture, local overstretching, open cell impingement, or stress concentration.As a result, although no withdrawal difficulties were reported, the applied pull force to remove the device could have then led to the reported separation.In addition, the technical summary outlines the extensive manufacturing mitigations in place to prevent this type of malfunction (visual and dimensional inspections, leak testing, and functional balloon burst testing that occurs with every manufactured lot).These inspections and testing further support that it is unlikely that a defect present in manufacturing contributed to the complaint.The technical summary also outlines the instructions for valve deployment.It should be noted that these mitigations are still in place.Review of available information suggests that patient factors (calcification ) contributed to the balloon burst while procedural factors (excessive manipulation) contributed to the separation.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
|
|
Search Alerts/Recalls
|
|
|