Model Number HL-90 |
Device Problem
Power Problem (3010)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the fluid warmers power cord is damage.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
|
|
Manufacturer Narrative
|
D10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Event problem and evaluation codes: updated.One device was received.The physical condition noted a cracked tank cover, broken pole clamp and damaged line cord.There is a cut in the outer sheath of the power cord confirming the customer complaint.Root cause was determined to be damage due to wear/usage.A manufacturing device history record (dhr) review was not performed because the device is beyond a year from its manufacture date of 2000 and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.
|
|