• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Moisture or Humidity Problem (2986)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2020
Event Type  malfunction  
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) formed condensation in the tubing.No patient harm, serious injury or adverse event was reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported "condensation" issue.However, the fse found brownish blackish material leading into the tidal disk.To fix the issue, the fse replaced the pim, from customer's stock.The iabp was able to pass all the leak tests, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.This report is being submitted as the result of a retrospective review conducted in capa (b)(4).
 
Manufacturer Narrative
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) formed condensation in the tubing.A getinge field service engineer evaluated found brownish blackish material leading into the tidal disk.Replaced pneumatic interface module (pim).Pim from customers stock.Performed full functional and safety tests.All tests pass.Returned to customer and cleared for clinical use.There was a patient involvement but no harm reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16269192
MDR Text Key308758388
Report Number2249723-2023-00808
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2018
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age71 YR
Patient SexFemale
Patient Weight40 KG
-
-