DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Moisture or Humidity Problem (2986)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/03/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) formed condensation in the tubing.No patient harm, serious injury or adverse event was reported.
|
|
Manufacturer Narrative
|
A getinge field service engineer (fse) evaluated the iabp and was unable to reproduce the reported "condensation" issue.However, the fse found brownish blackish material leading into the tidal disk.To fix the issue, the fse replaced the pim, from customer's stock.The iabp was able to pass all the leak tests, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.This report is being submitted as the result of a retrospective review conducted in capa (b)(4).
|
|
Manufacturer Narrative
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) formed condensation in the tubing.A getinge field service engineer evaluated found brownish blackish material leading into the tidal disk.Replaced pneumatic interface module (pim).Pim from customers stock.Performed full functional and safety tests.All tests pass.Returned to customer and cleared for clinical use.There was a patient involvement but no harm reported.
|
|