• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/19/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented in the operation room for implant procedure.During the procedure, the newly implanted left ventricular (lv) lead was found damaged, and portions of the lead was separated from the main body of the lead.The left ventricular lead was not installed and the procedure was completed using an alternate left ventricular lead on (b)(6) 2023.The patient's condition was stable.
 
Manufacturer Narrative
The reported event of ¿the lead is damaged and part of the lead separated from the main body of the lead¿ was confirmed.As received, a partial lead was returned in one piece measuring 86.0 cm.It was noted that the connector pin and connector cap of the lead were pulled out of the molded connector stretching and separating the inner coil of the lead which consistent with damage due to excessive forces applied during the procedure.The connector pin with the cap were missing/not returned with the lead.The cause of the reported event was consistent with procedural damage.A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16270364
MDR Text Key308454564
Report Number2017865-2023-04587
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeCH
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000126360
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/03/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-